The resolution index for command-response and AIN-EF readings. A larger resolution index generally results in lower noise and longer sample times.
- Data type: UINT16 (type index = 0)
- Readable and writable
- Default value: 0
-
T8:
- Valid values: 0-16. A value of 0 will instruct the T8 to use the best resolution for the rate specified.
-
T7:
- Valid values: 0-8 for T7, 0-12 for T7-Pro. Default value of 0 corresponds to an index of 8 (T7) or 9 (T7-Pro).
-
T4:
- Valid values: 0-5. Default value of 0 corresponds to an index of 5.
Expanded Names | Addresses |
AIN0_RESOLUTION_INDEX, AIN1_RESOLUTION_INDEX, AIN2_RESOLUTION_INDEX, AIN3_RESOLUTION_INDEX, AIN4_RESOLUTION_INDEX, AIN5_RESOLUTION_INDEX, AIN6_RESOLUTION_INDEX, AIN7_RESOLUTION_INDEX, AIN8_RESOLUTION_INDEX, AIN9_RESOLUTION_INDEX, AIN10_RESOLUTION_INDEX, AIN11_RESOLUTION_INDEX, AIN12_RESOLUTION_INDEX, AIN13_RESOLUTION_INDEX Show All | 41500, 41501, 41502, 41503, 41504, 41505, 41506, 41507, 41508, 41509, 41510, 41511, 41512, 41513 Show All |